Oh, Park, Baek, and Woo: Application of the Bair Hugger™ core body temperature at wrist region with upper body warming blanket: a prospective observational study

Abstract

Background

Body temperature monitoring is essential during the perioperative period. However, core body temperature measurement requires invasive device that may cause complications. This study aimed to evaluate the accuracy of non-invasive Bair Hugger™ core body temperature monitoring system (BHTMS) at the wrist compared with esophageal temperature under general anesthesia.

Methods

Twenty adult patients of the American Society of Anesthesiologists physical status I or II were enrolled. BHTMS sensor was applied at wrist region. After tracheal intubation, an esophageal probe was inserted. Bair Hugger™ upper body warming blankets were used. Esophageal temperature (Teso) and BHTMS at wrist (Twrist) were recorded every 10 min.

Results

Total of 257 pairs of data sets were analyzed: Teso and Twrist had no statistically significant difference (P = 0.103). Median of Teso and Twrist were 36.5℃ and 36.4℃. Bland-Altman analysis showed Teso - Twrist of 0.14℃ ± 1.44. Subsequently, 99 pairs of 0-40 min data set were analyzed and showed significant difference at 0 and 10 min (P < 0.001) but no significant difference at 20, 30 and 40 min. Bland-Altman plot by times showed difference (Teso - Twrist) of 1.49℃ ± 2.00, 0.82℃ ± 1.30, 0.29℃ ± 1.32, -0.03℃ ± 0.84, and -0.12℃ ± 0.82 at 0, 10, 20, 30 and 40 min respectively.

Conclusions

BHTMS at wrist area under the upper body warming blanket is a potential alternative other than esophageal temperature for monitoring body temperature after 30 min of anesthesia induction.

INTRODUCTION

Body temperature monitoring during the perioperative period is essential because of the variability during surgery. Body temperature often decreases during surgery because of thermoregulation disruption due to general anesthesia and cold exposure to procedural surroundings [1-4]. Perioperative hypothermia may cause mild consequences such as patient’s discomfort, shivering and severe complications such as increased bleeding [1] due to abnormal coagulopathy [2], higher risk of surgical site infection [3] and life-threatening arrhythmia [4]. Meanwhile, body temperature may increase soon after the induction of general anesthesia if malignant hyperthermia is triggered by exposure to volatile anesthetic agents or succinylcholine [5]. Immediate recognition and treatment should be started without delay in the occurrence of malignant hyperthermia [6]. Therefore, we should be able to monitor perioperative body temperature accurately.
Nevertheless, skin temperature may differ greatly from the actual body temperature; thus core body temperature must be measured to ensure accuracy [7]. Core body temperature can be obtained from the tympanic membrane, nasopharynx, esophagus, pulmonary artery, bladder and rectum. However, temperature measurements at these sites require using invasive placement of sensors.
Esophageal temperature is strongly associated with pulmonary artery temperature and is used as a basic temperature measurement method for intubated patients [8]. Esophageal probe insertion is usually considered to be a safe method; however, there is the possibility of complications such as esophageal bleeding, perforation, and arrhythmias [8]. Moreover, inserting esophageal probe is difficult when supraglottic airway is used.
To measure core body temperature in a non-invasive way, Fox and Solman [9] first invented the non-invasive zero-heat-flux thermometer in 1971. 3M developed a 3M™ Bair Hugger™ temperature monitoring instrument for use in clinical practice. It is a zero-heat-flux thermometer consisting of two thermistors, a thermal insulator and an electrical heater. A sensor is put on the forehead region and insulates the skin locally by heating. An area of isothermic tunnel is then created from the skin to the core body part such as the brain. At that point, the state of zero heat flux reflects the core body temperature [10-12]. Safety and usefulness of this method have already been reported. Previous studies have proved the reliability of this method through comparison with esophageal probe [10], nasopharyngeal probe [11], and pulmonary artery catheter [13]. The Bair Hugger™ temperature monitoring system (BHTMS) has some advantages, such as its accuracy, ease of use with disposable sensors, and non-invasiveness. However, the BHTMS may not be able to be applied to the forehead area owing to several reasons such as the field of surgery involving the head, skin problem at the attachment site, and interference of other monitoring probes on the forehead.
Therefore, instead of the forehead area, the BHTMS sensor is applied to another area of the body. According to Tachibana et al. [14] a relatively accurate temperature could be obtained even if the BHTMS sensor was attached to the neck. However, temperature obtained at the chest regions did not show accurate values. Considering the principle of zero-heat-flux thermometer, we hypothesized that core body temperature can be obtained from the wrist area using BHTMS. The wrist area has a radial artery running about 5 mm below the skin, has little influence by fatty tissues, and is easy to approach [15]. To date, no studies have focused on the use of BHTMS at the wrist region. Hence this study aimed to determine whether an accurate core body temperature can be obtained from the wrist region using BHTMS in patients under general anesthesia by comparing it with esophageal temperature.

MATERIALS AND METHODS

This study was approved by the Institutional Review Board of Chung-Ang University Hospital (registration no: 2112F-030-489) and it was registered at the Clinical Research Information Service clinical trials registry (KCT0007211) before patient recruitment. Written informed consent was obtained from each patient before surgery.
The study was conducted between March 2022 and August 2022. Patients of the American Society of Anesthesiologists physical status I or II who were planned to undergo orthopedic surgery under general anesthesia were included. Pediatric patients, pregnant patients, patients with esophageal lesions such as esophageal varix or stricture, abnormalities at the wrist region, and hematologic disease of bleeding tendency were excluded. We included patients with minimal risk of massive bleeding during surgery.
Before the induction of general anesthesia, the BHTMS sensor (3M™ Bair Hugger™ Temperature Monitoring Patient Sensor, 36000 and 3M™ Bair Hugger™ Temperature Monitoring System, 37000) was attached to the patient’s wrist. The center of the sensor was placed at the radial artery pulsation site. Anesthesia was induced by propofol (1.5-2.0 mg/kg) and remifentanil (0.5 μg/kg), and tracheal intubation was performed 90 s after injecting 0.6-0.9 mg/kg rocuronium. Maintenance of anesthesia was performed using air-oxygen-sevoflurane with FiO2 0.5. Sevoflurane dosage was adjusted at 1.5-2.0 minimum alveolar concentration to maintain patient state index of 25-50 at the SEDLine monitor (SEDLine™, Masimo, CA). After the tracheal intubation, an esophageal probe (ETP1040, Ewha Biomedics) was inserted through the mouth with the insertion depth determined using the equation (depth [cm] = 0.228 × standing height [cm] - 0.194) to target the region of esophagus bounded by the left ventricle and aorta [16]. Temperature recording began after equilibration, which was defined as 0 min [T0]. Patients were warmed appropriately at the upper body using a Bair Hugger™ forced-air warming unit (3M™ Bair Hugger™ Warming Unit Model 775, 3M) and Bair Hugger™ upper body warming blankets (3M™ Bair Hugger™ Warming Blanket Model 622, Multi-position, 3M), under high operating temperature (43℃). Warming blanket covered the whole upper body including the wrist area. The warming system was turned off when esophageal temperature (Teso) reached over 37.3℃. Teso and BHTMS at the wrist region (Twrist) were recorded every 10 min until the end of the surgery. Before extubation, esophageal temperature probe and BHTMS sensor were removed. The operating room temperature was maintained at 20-24℃ throughout the study period.
Our primary end point was comparison of consistency between Teso and Twrist. The secondary end points include the accuracy and correlation between Teso and Twrist by times at 0, 10, 20, 30, and 40 min (T0, T10, T20, T30, and T40, respectively), and accuracy between Teso and Twrist after 30 min of anesthesia induction.

Statistical analysis

During the research planning, we performed a pilot study, collecting 132 pairs of data from 11 patients. After the pilot study, the sample size was calculated with G Power software (Ver. 3.1.9.7, Heinrich-Heine-Universität Düsseldorf) using a paired t-test. The mean difference was 0.15 and the standard deviation (SD) was 0.38. When calculated with an alpha value of 0.01 and power of 0.8, the sample size was 79. In the pilot study, the shortest duration of surgery was 40 min. It means 1 patient yielded a minimum of 5 data sets. Therefore, we divided 79 by 5 to calculate minimum number of participants. Then, considering 20% of drop out rate, 20 patients were decided as the total number of participants.
The distribution of parameters was evaluated for normality using the Kolmogorov-Smirnov test. Non-normally distributed data were compared using the Wilcoxon signed rank test and Spearman correlation and are expressed as median and interquartile range. Normally distributed data were compared using the paired t-test and Pearson correlation and are expressed as mean and SD. For the comparison of Teso and Twrist data by times, Bonferroni correction was performed. P value < 0.01 was considered significant. Bland-Altman plots were also used to evaluate the limits of agreement and were expressed as mean bias (Teso - Twrist) ± 2SD. All statistical analyses were performed using dBSTAT for Windows (Ver. 5.0, dBSTAT).

RESULTS

A total 258 pairs of data were collected from 20 patients. Table 1 summarizes the demographic data and duration of anesthesia. In one case, a pair of data at 0 min was deleted because of Twrist missing data due to an electronical problem. Therefore, 257 pairs of data were analyzed using the Wilcoxon signed rank test and Spearman correlation. Median (1Q, 3Q) of Teso and Twrist were 36.5℃ (35.95℃, 36.8℃) and 36.4℃ (35.8℃, 36.8℃), respectively. Teso and Twrist had no statistically significant difference (P = 0.103) (Fig. 1). The Spearman’s correlation coefficient was 0.567 (Fig. 2).
Each patient had a different length of anesthesia time according to the type of surgery. Table 2 summarizes the types of surgery. The shortest surgery duration was 40 min. Overall, 99 pairs of data were organized and analyzed according to the measurement time intervals (0, 10, 20, 30 and 40 min). The paired t-test and Pearson correlation were performed, and Table 3 summarizes the result of paired t-test. A significant difference was observed at T0, T10 and T20 whereas no significant difference was observed at T30 and T40. Fig. 3 shows the Pearson’s correlation coefficients of Teso and Twrist by times, which were 0.262, 0.606, 0.679, 0.746, and 0.718 at T0, T10, T20, T30, and T40 respectively.
A Bland-Altman analysis was also performed. For the primary outcome, mean difference (Teso - Twrist) of 257 pairs of data was 0.14℃ with a 2SD of ± 1.44 (Fig. 4). For the secondary outcome, the Bland-Altman plot was constructed for Teso and Twrist by times. The mean bias (Teso - Twrist) ± 2SD were 1.49℃ ± 2.00, 0.82℃ ± 1.30, 0.29℃ ± 1.32, -0.03℃ ± 0.84, -0.12℃ ± 0.82 at T0, T10, T20, T30, and T40, respectively (Fig. 5).

DISCUSSION

This study aimed to determine whether an accurate core body temperature can be obtained from the wrist region using BHTMS in patients under general anesthesia by comparing it with esophageal temperature. We compared Teso and Twrist as the primary outcome. Wilcoxon signed-rank test showed that Teso and Twrist had no significant difference. Spearman correlation showed a moderate positive correlation. However, the Bland-Altman plot of Teso and Twrist showed a mean bias (Teso - Twrist) ± 2SD of 0.14 ± 1.44℃.
Boisson et al. [10] compared the temperature recorded using BHTMS at the forehead with esophageal temperature, and the result showed mean bias and limit of agreement (LOA) of 0.1 ± 0.5℃. Iden et al. [11] compared the temperature recorded using this method with nasopharyngeal temperature, showing almost identical results (0.07 ± 0.42℃). Eshraghi et al. [13] compared the temperature recorded using BHTMS with pulmonary artery temperature and found a result of -0.23 ± 0.82℃. Also, Tachibana et al. [14] compared the temperature recorded using BHTMS at the neck with esophageal temperature, showing a result of 0.05 ± 0.35℃. Tachibana et al. [14] considered a mean value of the difference (bias) < 0.4℃ and 2SD < ± 1.0℃ as appropriate standard for comparing the two methods. In this study, we compared Teso and Twrist by considering the same standard as Tachibana’s study and the result of mean bias (Teso - Twrist) ± 2SD of 0.14 ± 1.44℃ did not meet the agreed standard.
For the secondary outcomes, the T0, T10, T20, T30, and T40 data were analyzed. T0 and T10 showed significant differences in the paired t-test, whereas T20, T30, and T40 showed no significant difference. Additionally, there was a positive correlation between Teso and Twrist at T0-T40, although the T0 showed a weak correlation. However, T10, T20, T30, and T40 showed moderate to strong correlations. From the Bland-Altman analysis result, we noticed that only T30 and T40 could be accepted as reliable, which met the earlier LOA standard [14]. For further analysis, we also performed Bland-Altman analysis with data collected from 30 min to the end of the surgery. It showed a mean bias (Teso - Twrist) ± 2SD of 0.08 ± 0.96℃, which met the agreed standard. Considering these results, temperature measurement using the BHTMS at the wrist is relatively accurate and equivalent to the esophageal temperature at least 30 min after the first temperature measurement.
In this study, at T0, 10 of 20 patients showed immediate close approximation between Teso and Twrist ≤ ± 1.0℃. However, the remaining half of patients showed a wide difference between Teso and Twrist, up to 3.1℃, and the wide difference decreased progressively. The exact mechanism of the close approximation of Twrist to Teso after 30 min needs to be explicit. One reason, we assume, is the redistribution of body heat. Before the induction of general anesthesia, the peripheral compartment distant from the heart, such as the wrist region, typically shows a 2-4℃ lower temperature than the core body temperature. This is the normal core-to-peripheral temperature gradient. General anesthesia is known to reduce vasoconstriction and cause peripheral vasodilation, which leads to perioperative hypothermia. The following redistribution increases the temperature of the peripheral region [17]. Furthermore, we assumed that the other reason is the direct heating effect of the upper body warming blanket. When we turned off the warming system owing to concern of hyperthermia, we could observe a wider difference between Teso and Twrist. Therefore, we believed that the upper body warming blanket has a direct heating effect. The time interval for the calibration of the BHTMS sensor and the time interval until the application of a warming blanket is another factor considered in the direct heating effect. Considering the calibration time, we applied the BHTMS sensor before intubation. Calibration was done within 5 min in every patient before placing the esophageal temperature probe. A warming blanket was applied after intubation or after positioning the patients for the operation. The time interval was not correctly measured during the study; however, it might have been up to 15 min.
Our study had several limitations. First, we monitored and analyzed a esophageal temperature range of 35.2-37.2℃ in patients undergoing orthopedic surgeries. Therefore, we could not assure that BHTMS would correctly measure hyperthermia or hypothermia out of that range or in other surgery types. Second, for the secondary outcome, we did not compare the temperature by times after 40 min because the surgery time length differed in every patient. Thus, we could not perform comparisons of data by times after 40 min. Third, we did not include patients who did not use the warming system in the study because of the potential ethical issues. Also, we did not use warming system other than the upper body warming blanket. Further study using warming system other than the upper body warming blanket is needed. Fourth, we did not measure the BHTMS temperature at the forehead region and the peripheral skin temperature for comparison. Further studies comparing temperatures recorded using the BHTMS at the forehead, wrist, and peripheral skin temperature is needed. Fifth, the peripheral circulation of each patient was not evaluated although it might be concerned whether the location of radial artery was temperature of the fluid were not evaluated. However, considering the result of the secondary outcome, under the Bair Hugger™ upper body warming blankets, the influence of fluid after 30 min could be considered minimal.
Limitations stated above may lower the strength of this study. However, this study has meaningful number of participants and well documented methodology to be replicated. Also, Ethical guidelines were followed and potential threat of hypothermia or hyperthermia was prevented. Further studies should be proceeded with larger number of participants for longer duration of time, overall control of the potential variables that may affect body temperature, and comparison with other temperature sites. In this study, results were inconclusive about the accuracy of BHTMS at wrist area. Even though, this study has significance for the suggestion of another possible alternative method for monitoring body temperature.
In conclusion, the accuracy of BHTMS at wrist area under upper body warming blanket is comparable to that of esophageal temperature after 30 min of anesthesia induction. Therefore, BHTMS may be another possible alternative method for monitoring body temperature after 30 min of anesthesia induction.

Notes

FUNDING

None.

CONFLICTS OF INTEREST

No potential conflict of interest relevant to this article was reported.

DATA AVAILABILITY STATEMENT

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

AUTHOR CONTRIBUTIONS

Writing - original draft: Kyung Seo Oh. Writing - review & editing: Kyung Seo Oh, Young-Cheol Woo. Conceptualization: Young-Cheol Woo. Methodology: Young-Cheol Woo. Project administration: Kyung Seo Oh. Investigation: Kyung Seo Oh. Supervision: Yong-Hee Park, Chongwha Baek, Young-Cheol Woo.

Fig. 1.
Comparison of Teso and Twrist. Wilcoxon signed rank test was performed, and no statistical difference was observed (P = 0.103). Teso: esophageal temperature, Twrist: temperature recorded using the Bair Hugger™ core body temperature monitoring system at the wrist.
apm-23107f1.jpg
Fig. 2.
Spearman correlation between Teso and Twrist. Teso: esophageal temperature, Twrist: temperature recorded using the Bair Hugger™ core body temperature monitoring system at the wrist, Rs: Spearman correlation coefficient.
apm-23107f2.jpg
Fig. 3.
Pearson correlation between Teso and Twrist by times. Teso and Twrist were recorded at 0 min (A), 10 min (B), 20 min (C), 30 min (D) and 40 min (E). Teso: esophageal temperature, Twrist: temperature recorded using the Bair Hugger™ core body temperature monitoring system at the wrist, R: Pearson correlation coefficient.
apm-23107f3.jpg
Fig. 4.
Bland-Altman plot of Teso and Twrist. Teso: esophageal temperature, Twrist: temperature recorded using the Bair Hugger™ core body temperature monitoring system at the wrist, SD: standard deviation.
apm-23107f4.jpg
Fig. 5.
Bland-Altman plots of Teso and Twrist at 0 min (A), 10 min (B), 20 min (C), 30 min (D), and 40 min (E). Teso: esophageal temperature, Twrist: temperature recorded using the Bair Hugger™ core body temperature monitoring system at the wrist, SD: standard deviation.
apm-23107f5.jpg
Table 1.
Demographic Data
Variable Value (n=20)
Sex (M / F) 11/9
Age (yr) 47.8± 19.3
Height (cm) 165.8± 8.8
Weight (kg) 69.9± 12.8
BMI (kg/m2) 25.3± 3.2
Anesthesia time (min) 129± 52.7
ASA-PS (I/II) 10/10

Values are presented as number only or mean ± SD. BMI: body mass index, ASA PS: American Society of Anesthesiologists physical status.

Table 2.
Types of Surgery
Arthroscopic surgery of knee or shoulder 3
Total knee replacement 4
Metal removal of ankle or knee or clavicle 6
ORIF of ankle or hand or humerus 3
Ligament reconstruction of elbow or knee 2
Ulnar osteotomy 2
Total 20

Values are presented as number only. ORIF: open reduction and internal fixation

Table 3.
Comparison of Teso and Twrist (℃) by Times
Teso Twrist P value
0 min 36.31 ± 0.54 34.82 ± 0.99 < 0.001
10 min 36.18 ± 0.53 35.36 ± 0.82 < 0.001
20 min 36.18 ± 0.50 35.89 ± 0.88 0.063
30 min 36.21 ± 0.53 36.24 ± 0.62 0.751
40 min 36.25 ± 0.52 36.38 ± 0.56 0.203

Values are presented as mean ± SD. Teso: esophageal temperature, Twrist: temperature recorded using the Bair Hugger™ core body temperature monitoring system at the wrist.

The two groups were compared using the paired t-test. Bonferroni correction was performed on the data. P value < 0.01 was considered significant.

REFERENCES

1. Schmied H, Kurz A, Sessler DI, Kozek S, Reiter A. Mild hypothermia increases blood loss and transfusion requirements during total hip arthroplasty. Lancet 1996; 347: 289-92.
crossref pmid
2. Rohrer MJ, Natale AM. Effect of hypothermia on the coagulation cascade. Crit Care Med 1992; 20: 1402-5.
crossref pmid
3. Kurz A, Sessler DI, Lenhardt R. Perioperative normothermia to reduce the incidence of surgical-wound infection and shorten hospitalization. Study of Wound Infection and Temperature Group. N Engl J Med 1996; 334: 1209-15.
crossref pmid
4. Frank SM, Fleisher LA, Breslow MJ, Higgins MS, Olson KF, Kelly S, et al. Perioperative maintenance of normothermia reduces the incidence of morbid cardiac events. A randomized clinical trial. JAMA 1997; 277: 1127-34.
crossref pmid
5. Hopkins PM. Malignant hyperthermia: advances in clinical management and diagnosis. Br J Anaesth 2000; 85: 118-28.
crossref pmid
6. Rosenberg H, Davis M, James D, Pollock N, Stowell K. Malignant hyperthermia. Orphanet J Rare Dis 2007; 2: 21.
crossref pmid pmc pdf
7. Severinghaus JW. Temperature gradients during hypothermia. Ann N Y Acad Sci 1959; 80: 515-21.
crossref pmid
8. Hymczak H, Golab A, Mendrala K, Plicner D, Darocha T, Podsiadlo P, et al. Core temperature measurement-principles of correct measurement, problems, and complications. Int J Environ Res Public Health 2021; 18: 10606.
crossref pmid pmc
9. Fox RH, Solman AJ. A new technique for monitoring the deep body temperature in man from the intact skin surface. J Physiol 1971; 212: 8-10.

10. Boisson M, Alaux A, Kerforne T, Mimoz O, Debaene B, Dahyot-Fizelier C, et al. Intra-operative cutaneous temperature monitoring with zero-heat-flux technique (3M SpotOn) in comparison with oesophageal and arterial temperature: a prospective observational study. Eur J Anaesthesiol 2018; 35: 825-30.
pmid
11. Iden T, Horn EP, Bein B, Böhm R, Beese J, Höcker J. Intraoperative temperature monitoring with zero heat flux technology (3M SpotOn sensor) in comparison with sublingual and nasopharyngeal temperature: an observational study. Eur J Anaesthesiol 2015; 32: 387-91.
pmid
12. Pesonen E, Silvasti-Lundell M, Niemi TT, Kivisaari R, Hernesniemi J, Mäkinen MT. The focus of temperature monitoring with zero-heat-flux technology (3M Bair-Hugger): a clinical study with patients undergoing craniotomy. J Clin Monit Comput 2019; 33: 917-23.
crossref pmid pdf
13. Eshraghi Y, Nasr V, Parra-Sanchez I, Van Duren A, Botham M, Santoscoy T, et al. An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients. Anesth Analg 2014; 119: 543-9.
crossref pmid
14. Tachibana S, Chida Y, Yamakage M. Using the Bair HuggerTM temperature monitoring system in neck and chest regions: a pilot study. JA Clin Rep 2019; 5: 32.
pmid pmc
15. Kang IG, Jeong WJ, Moon KM. The relationship between radial artery depth and wrist extension angle measured by ultrasonography. J Korean Soc Emerg Med 2013; 24: 279-83.

16. Mekjavić IB, Rempel ME. Determination of esophageal probe insertion length based on standing and sitting height. J Appl Physiol (1985) 1990; 69: 376-9.
crossref pmid
17. Sessler DI. Perioperative heat balance. Anesthesiology 2000; 92: 578-96.
crossref pmid pdf


ABOUT
ARTICLE & TOPICS
Article category

Browse all articles >

Topics

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: apm@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next