Interruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia: A case report
Hyea Hyoung Cho, Sung Uk Choi, Ji Yong Park, Hye Won Shin
Department of Anesthesiology and Pain Medicine, Korea University Medical Center, Anam Hospital, Seoul, Korea. drchois@korea.ac.kr
Abstract
We report our experience of unexpected low expiratory concentration of sevoflurane 0.8 vol%, low inspired fraction of oxygen 31%, and high value of bispctral index (BIS) 62 with sevoflurlane vaporizer setting 2.0 vol% and the inspired fraction of oxygen 50% during general anesthesia of a 56-year-old man undergoing elective laparoscopic hemicolectomy. The anesthesia machine (Cato edition, Drager, Lubeck, Germany) with a piston-based, electronically controlled ventilator was used for anesthesia. We checked thoroughly all the possible causes of the discrepancy of anesthetic gas concentration, and then detected the partial disconnection of fresh gas outlet. But there were no alarm signs of anesthetic monitor with adequate ventilator function. This malfunction of anesthetic gas supply without alarm signs may lead to hypoxia and awareness of patients. The intraoperative monitoring of anesthetic gas analyzer and BIS may be essential to detect the malfunction of anesthesia machine.
Key Words: Alarm, Disconnection, Fresh gas outlet, Ventilator


ABOUT
ARTICLE & TOPICS
Article category

Browse all articles >

Topics

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: apm@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next